Equipment needed

• Sheet of stickers with measurement units (mm, inches and %)
• Small slotted tip screwdriver (not supplied)

Get access to the keypad on the indicator board

1 Remove the transmitter housing cover.
2 Remove the screws that hold the indicator board on the housing.
3 Turn the indicator board to use the 3-button keypad ([

], [F

F

F

F] and [

]) on the back of the board

to change the units. Refer to the tables that follow for key functions and the example proce-
dure for configuration of the indicator.

4 Attach the indicator board to the housing.
5 Put the sticker with the correct measurement unit over the default units under the indicator

screen.

6 Attach the transmitter housing cover.

Figure 6-3: Get access to the keypad on the indicator board