Menu overview
Example procedure: Change the indicator units from mm to %
• Push [F
F
F
F] for 3 seconds (button on the back of the indicator board) to access Program mode
Program mode
Program mode
Program mode.
i
In program mode dP
dP
dP
dP is displayed.
• Push [
] 1 time to get to the zero point (ZErO
ZErO
ZErO
ZErO) function.
• Push [F
F
F
F] to access the span range parameter. Make sure that the value is 0.0. Push [
] + [
]
(Escape) to go back to the menu.
• Push [
] 1 time to get to the span range (SPAn
SPAn
SPAn
SPAn) function.
• Push [F
F
F
F] to access the span range parameter.
i
The current value is 600 (example value).
• Push [
] to decrease the value to 100.
• Push [F
F
F
F] to confirm the parameter and go back to the menu.
• Push [
] + [
] (Escape) to go back to the indication.
i
The indication will display a value between 0 and 100 based on a current output of 4...20 mA.
• Put a new sticker (%) on the front of the indicator board.
Function
Function description
Range/Selection list
Default
Decimal Point (dP)
Push [
] or [
] to select
the decimal point position.
Push [F
F
F
F] to confirm.
0…3
1
Zero point (ZErO)
Push [
] or [
] to select
the minimum value. Push
[F
F
F
F] to confirm.
-1999…+9999
0.0
Span range (SPAn)
Push [
] or [
] to select
the maximum value. Push
[F
F
F
F] to confirm.
-1999…+9999
100.0
Input current limit (Li)
Do not use.
n/a
0
Filter (FiLt)
Do not use.
n/a
2
Resolution (riS)
Do not use.
n/a
1