Operation – KROHNE UFC 400 FOUNDATION FIELDBUS EN User Manual
Page 47
OPERATION
4
47
UFC 400
www.krohne.com
05/2013 - 4002813701 - AD FF UFC 400 R01 en
WEIGHT_2
WEIGHT_2
WEIGHT_2
WEIGHT_2
Weight 2
R/W
Sets the weight of path 2 (Fct. D2.7).
1.0
Settings:
0.0…1.0
WEIGHT_3
WEIGHT_3
WEIGHT_3
WEIGHT_3
Weight 3
R/W
Sets the weight of path 3 (Fct. D2.8).
1.0
Settings:
0.0…1.0
WINDOW_END
WINDOW_END
WINDOW_END
WINDOW_END
Window End
R/W
Window end (Fct. D1.3)
2000.0 µs
Settings:
15.0…3200.0 µs
WINDOW_START
WINDOW_START
WINDOW_START
WINDOW_START
Window Start
R/W
Window start (Fct. D1.2)
32.0 µs
Settings:
15.0…3200.0 µs
XD_ERROR
XD_ERROR
XD_ERROR
XD_ERROR
Transducer Error
One of the transducer error codes defined in the FF
Transducer Specifications in section "4.7 Block Alarm
Subcodes".
Uninitialized
Settings:
•
16: Unspecified error
•
17: General error
•
18: Calibration error
•
19: Configuration error
•
20: Electronics Failure
•
21: Mechanical Failure
•
22: I/O Failure
•
23: Data Integrity Error
•
24: Software Error
•
25: Algorithm Error
ZERO_
ZERO_
ZERO_
ZERO_
PATH_1
PATH_1
PATH_1
PATH_1
Zero Offs. - Path 1
R/W
Result of zero calibration for path 1 (Fct. D3.1.1).
0.0 ps
Settings:
-1000000…1000000 ps
ZERO_
ZERO_
ZERO_
ZERO_
PATH_2
PATH_2
PATH_2
PATH_2
Zero Offs. - Path 2
R/W
Result of zero calibration for path 2 (Fct. D3.1.2).
0.0 ps
Settings:
-1000000…1000000 ps
ZERO_
ZERO_
ZERO_
ZERO_
PATH_3
PATH_3
PATH_3
PATH_3
Zero Offs. - Path 3
R/W
Result of zero calibration for path 3 (Fct. D3.1.3).
0.0 ps
Settings:
-1000000…1000000 ps
Parameter
Parameter
Parameter
Parameter
DD name
Access
Description and settings
Initial Value