beautypg.com

Operation – KROHNE UFC 400 FOUNDATION FIELDBUS EN User Manual

Page 30

background image

4

OPERATION

30

UFC 400

www.krohne.com

05/2013 - 4002813701 - AD FF UFC 400 R01 en

FORMAT_2P_2L

FORMAT_2P_2L

FORMAT_2P_2L

FORMAT_2P_2L

Form. 2.Page 2.Line

R/W

No. of decimal places for 2st line of 2nd page (Fct. C5.4.4).

X.X

Setting:

#X. / #X.X / #X.XX / #X.XXX / #X.XXXX / #X.XXXXX /

#X.XXXXXX / #X.XXXXXXX / #X.XXXXXXXX / automatic

FORMAT_2P_3L

FORMAT_2P_3L

FORMAT_2P_3L

FORMAT_2P_3L

Form. 2.Page 3.Line

R/W

No. of decimal places for 3st line of 2nd page (Fct. C5.4.5).

X.X

Setting:

#X. / #X.X / #X.XX / #X.XXX / #X.XXXX / #X.XXXXX /

#X.XXXXXX / #X.XXXXXXX / #X.XXXXXXXX / automatic

IO_CONNECTION

IO_CONNECTION

IO_CONNECTION

IO_CONNECTION

Status IO Connection

R/W

Defines the influence of the diagnostic ''Electronic : IO

Connection'' (FD_xx_14) to the TB measurements status (Fct.

C1.10.7).

Information

Setting:

Information / Maintenance Required / Checks in progress /

Out Of Specification / Failure

PASSWRD

PASSWRD

PASSWRD

PASSWRD

Password

R/W

Password to activate access rights to restricted parameters.

0

POWER_FAILURE

POWER_FAILURE

POWER_FAILURE

POWER_FAILURE

Status Power Failure

R/W

Defines the influence of the diagnostic ''Electronic : Power

Failure'' (FD_xx_8) to the TB measurements status (Fct.

C1.10.8).

Information

Setting:

Information / Maintenance Required / Checks in progress /

Out Of Specification / Failure

PROTECT_INFO

PROTECT_INFO

PROTECT_INFO

PROTECT_INFO

Access protection

R

Current protection level to restricted parameters.

Production

Setting:

Concentration / bit 5…6: not use / User / Service level /

Production / Custody Transfer / Write Lock

SERIAL_NUMBER

SERIAL_NUMBER

SERIAL_NUMBER

SERIAL_NUMBER

Serial Number

R

Device serial number, serial number of the system.

Current serial

number

SIGNAL_LOST

SIGNAL_LOST

SIGNAL_LOST

SIGNAL_LOST

Status Signal Lost

R/W

Defines the influence of the diagnostic ''Process : Signal

Lost'' (FD_xx_12) to the TB measurements status (Fct.

C1.10.4).

Information

Setting:

Information / Maintenance Required / Checks in progress /

Out Of Specification / Failure

SIGNAL_UNRELIABLE

SIGNAL_UNRELIABLE

SIGNAL_UNRELIABLE

SIGNAL_UNRELIABLE

Status Signal

Unreliable

R/W

Defines the influence of the diagnostic ''Process : Signal

Unreliable'' (FD_xx_11) to the TB measurements status (Fct.

C1.10.5).

Information

Setting:

Information / Maintenance Required / Checks in progress /

Out Of Specification / Failure

SW_SIMULATE

SW_SIMULATE

SW_SIMULATE

SW_SIMULATE

SW simulate

R/W

Enable/disable simulation like simulate Hardware jumper.

Disable

Setting:

Disable / Active

Variable_2P_1L

Variable_2P_1L

Variable_2P_1L

Variable_2P_1L

Disp. 2.Page 1.Line

R/W

Select the Function Block output value to display at 2nd page

1st line (Fct. C5.4.1).

AI1

Setting:

AI10 / AI2 / AI3 / AI4 / PID / IT1 / IT2

Variable_2P_2L

Variable_2P_2L

Variable_2P_2L

Variable_2P_2L

Disp. 2.Page 2.Line

R/W

Select the Function Block output value to display at 2nd page

2nd line (Fct. C5.4.3).

AI1

Setting:

AI10 / AI2 / AI3 / AI4 / PID / IT1 / IT2

Parameter

Parameter

Parameter

Parameter
DD name

Access

Description and settings

Initial Value