(1)(11)(31):Check the acoustic properties of the medium for

this path / (1)(12)(29):Remove the damping or blockage in path

3 in the sensor. / (1)(12)(30):Remove the damping or blockage

in path 2 in the sensor. / (1)(12)(31):Remove the damping or

blockage in path 1 in the sensor. / (1)(13)(31):Fill the sensor

with liquid to return to normal operation. / (1)(14)(28):Check

cable and connection at terminal B , reduce load. /

(1)(14)(29):Check cable and connection at terminal B , reduce

load. / (1)(14)(30):Check cable and connection at terminal A ,

reduce load. / (1)(14)(31):Check cable and connection at

terminal A , reduce load. / (1)(15)(30):Check device connected

to current input B. / (1)(15)(31):Check device connected to

current input A. / (1)(16)(00):reserved / (1)(17)(30):Save

settings to Backup 2. / (1)(17)(31):Save settings to Backup 1. /

(1)(18)(29):Change one param. The status should disappear

within 1 minute. / (1)(18)(30):Please contact the factory. /

(1)(18)(31):Check if converter is correctly mounted and fixed

into housing. / (1)(19)(00):reserved / (1)(20)(31):Check process

conditions or pipe diameter settings. / (1)(21)(03):Check the

range settings of the related output. / (1)(21)(04):Check the

range settings of the related output. / (1)(21)(05):Check the

range settings of the related output. / (1)(21)(06):Check the

range settings of the related output. / (1)(21)(07):Check the

range settings of the related output. / (1)(21)(09):Check the

range settings of the related output. / (1)(22)(29):Recalibrate

FrontEnd of the converter or contact the factory. /

(1)(23)(00):reserved / (1)(24)(00):reserved / (1)(25)(06):Turn off

simulation / (1)(25)(07):Turn off simulation / (1)(25)(08):Turn

off simulation / (1)(25)(09):Turn off simulation /

(1)(25)(10):Turn off simulation / (1)(25)(11):Turn off

simulation / (1)(25)(12):Turn off simulation / (1)(25)(30):For

normal operation turn off simulation mode. / (1)(25)(31):For

normal operation turn off simulation mode. /

(1)(26)(00):reserved / (1)(27)(00):reserved /

(1)(28)(00):reserved / (1)(29)(02):Check IO B configuration or

reset to factory configuration. / (1)(29)(03):Check IO B

configuration or reset to factory configuration. /

(1)(29)(04):Check IO A configuration or reset to factory

configuration. / (1)(29)(05):Check IO A configuration or reset to

factory configuration. / (1)(29)(10):Check IO2 configuration or

reset to factory configuration. / (1)(29)(11):Check IO1

configuration or reset to factory configuration. /

(1)(29)(12):Check display configuration or reset to factory

configuration. / (1)(29)(21): Check fieldbus configuration or

reset to factory configuration / (1)(29)(26):Check config. of the

related function. Reset config. to factory. / (1)(29)(30):Check

config. of the related function. Reset config. to factory. /

(1)(29)(31):Check config. of the related function. Reset config.

to factory. / (1)(30)(01):Perform power reset. If the status

returns , contact the factory. / (1)(30)(02):Perform power

reset. If the status returns , contact the factory. /

(1)(30)(07):Perform power reset. If the status returns , contact

the factory. / (1)(30)(08):Perform power reset. If the status

returns , contact the factory. / (1)(30)(16):Perform power

reset. If the status returns , contact the factory. /

(1)(30)(18):Perform power reset. If the status returns , contact

the factory. / (1)(30)(19):Perform power reset. If the status

returns , contact the factory. / (1)(30)(20):Perform power

reset. If the status returns , contact the factory. /

(1)(30)(21):Perform power reset. If the status returns , contact

the factory. / (1)(30)(26):Perform power reset. If the status

returns , contact the factory. / (1)(30)(27):Perform power

reset. If the status returns , contact the factory. /

(1)(30)(28):Perform power reset. If the status returns , contact

the factory. / (1)(30)(31):Perform power reset. If the status

returns , contact the factory. / (1)(31)(31):Check manual for

transducer cabling in case of a remote device.