B.44 [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude]? QUERY

CURR:TRIG?

Syntax:

Short Form: [SOUR:]CURR[:LEV]:TRIG[:AMP]?
Long Form: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude]?
Return Value: = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Returns the current value established by CURR:TRIG command.

B.45 [SOURce:]FUNCtion:MODE COMMAND

FUNC:MODE

Syntax:

Short Form: FUNC:MODE {VOLT | CURR | EXT}
Long Form: [SOURce:]FUNCtion:MODE {VOLT | CURR | EXTernal}

Description: Establishes the operating mode of the power supply. VOLT = Constant Voltage mode (CV). CURR

= Constant Current mode (CC). EXT = External. FUNC:MODE VOLT commands power supply to Volt-
age mode, FUNC:MODE CURR commands power supply to Current mode. Commanded mode estab-
lishes parameters (voltage or current) monitored for error conditions. Actual mode depends upon load
conditions. When commanded to Voltage mode, if load conditions cause the power supply to try to
exceed the current limit, the unit will flag an error condition. When commanded to Current mode, if
load conditions cause the power supply to try to exceed the voltage limit, the unit will flag an error con-
dition.

B.46 [SOURce:]FUNCtion:MODE? QUERY

FUNC:MODE?

Syntax:

Short Form: FUNC:MODE?

Long Form: [SOURce:]FUNCtion:MODE?

Return value: 0 (voltage mode) or 1 (current mode)

Description: Returns the operating mode of the power supply. 0 = Constant Voltage mode (CV). 1 = Constant

Current mode (CC).

B.47 [SOURce:]FUNCtion:MODE:TRIGger COMMAND

FUNC:MODE:TRIG

Syntax:

Short Form: FUNC:MODE:TRIG {VOLT | CURR}
Long Form: [SOURce:]FUNCtion:MODE:TRIGger {VOLT | CURR}

Description: Establishes the operating mode of the power supply when a TRIGger command is sent. VOLT =

Constant Voltage mode (CV). CURR = Constant Current mode (CC). FUNC:MODE:TRIG VOLT com-
mands power supply to Voltage mode when a TRIGger command is sent, FUNC:MODE:TRIG CURR
commands power supply to Current mode when a TRIGger command is sent.

B.48 [SOURce:]FUNCtion:MODE:TRIGger? QUERY

FUNC:MODE:TRIG?

Syntax:

Short Form: FUNC:MODE:TRIG?

Long Form: [SOURce:]FUNCtion:MODE:TRIGger?

Return value: 0 = voltage mode or 1 = current mode

Description: Returns the operating mode which the power supply will enter when a trigger command is

sent. VOLT = Constant Voltage mode (CV). CURR = Constant Current mode (CC).

B.49 [SOURce:]LIST:CLEar COMMAND

LIST:CLE

Syntax:

Short Form: LIST:CLE

Long Form: LIST:CLEar>

Description: Clears all list entries by setting all pointers to 0. Releases control of the Output Off flag at EXT

PROTECT PORT, pins 5 and 6 (see Table 2-4), from the programmed list. Related Commands: All
LIST commands (See example, Figure B-5.)