4-6

BOP-1K 031014

2. Set the BOP to zero volts output by sending

CAL:VOLT ZERO

. Connect a Digital Voltmeter

(DVM) to the BOP OUT S and COM S terminals to measure the output voltage. Send

CAL:DATA

commands as needed (see PAR. 4.3a) to adjust the BOP output until the DVM

reads as close to zero as possible within tolerance specified in Table 4-3 for VOLTAGE
ZERO.

3. Set the BOP to maximum positive output voltage by sending

CAL:VOLT MAX

. Measure the

voltage output using the DVM. To adjust, send

CAL:DATA

commands as needed (see PAR.

4.3b) until the DVM reading is as close as possible above the nominal full scale value within
the tolerance specified in Table 4-3 for +FULL SCALE VOLTAGE.

4. Set the BOP to maximum negative output voltage by sending

CAL:VOLT MIN

. To adjust,

send the

CAL:DATA

command as needed (see PAR. 4.3b) until the reading is as close as

possible above (absolute value) the nominal full scale value within the limits specified in
Table 4-3 for –FULL SCALE VOLTAGE.

5. Send

CAL:VPR MAX

to adjust the maximum positive voltage protection limit of the power

supply. Send the

CAL:DATA

commands as needed (see PAR. 4.3c) until the reading is as

close as possible above the nominal full scale value within the limits specified in Table 4-3 for
+FULL SCALE VPR LIMIT.

6. Send

CAL:VPR MIN

to adjust the maximum negative protection limit of the power supply.

Send the

CAL:DATA

commands as needed (see PAR. 4.3c) until the reading is as close as

possible above (absolute value) the nominal full scale value within the limits specified in
Table 4-3 for –FULL SCALE VPR LIMIT.

7. Connect a 0V ±0.1mV d-c reference to pin 11 (EXT_REF) referenced to pin 10 (SGND) of

the Analog I/O Port connector (A2A5J6). Set the BOP to zero volts output by sending

CAL:VEXT ZERO

. Connect the DVM to the BOP output and send

CAL:DATA

commands as

needed (see PAR. 4.3a) to adjust the BOP output until the DVM reads as close to zero as
possible within tolerance specified in Table 4-3 for VOLTAGE ZERO.

NOTE: Accuracy of the 10V d-c reference must be ±0.1mV in order for the calibrated unit to
meet published specifications.

8. Replace 0V reference at pin 11 (EXT_REF) of the Analog I/O Port connector (A2A5J6) with a

+10V d-c reference. Set the BOP to maximum positive output voltage by sending

CAL:VEXT

MAX

. Measure the voltage output using the DVM. To adjust, send

CAL:DATA

commands as

needed (see PAR. 4.3b) until the DVM reading is as close as possible above the nominal full
scale value within the tolerance specified in Table 4-3 for +FULL SCALE VOLTAGE.

9. Replace +10V reference at pin 11 (EXT_REF) of the Analog I/O Port connector (A2A5J6)

with a 0V ±0.1mV d-c reference. Set the BOP to zero volts output by sending

CAL:VGA

ZERO

. Connect DVM to the BOP OUT S and OUT COM terminals to measure the output volt-

age. Send

CAL:DATA

commands as needed (see PAR. 4.3a) to adjust the BOP output until

the DVM reads as close to zero as possible within tolerance specified in Table 4-3 for VOLT-
AGE ZERO.

10.Replace 0V reference at pin 11 (EXT_REF) of the Analog I/O Port connector (A2A5J6) with a

+10V d-c reference. Set the BOP to maximum positive output voltage by sending

CAL:VGA

MAX

. Measure the output voltage using the DVM. To adjust, send

CAL:DATA

commands as

needed (see PAR. 4.3a) until the DVM reading is as close to zero as possible within toler-
ance specified in Table 4-3 for VOLTAGE ZERO.