3.3.12
ERROR MESSAGE EXPLANATIONS
When a key press is not accepted, an error message is displayed at the bottom of the screen.
Although recovery from most operator errors is obvious and simple, Table 3-14 lists all the error
messages along with associated explanations as to why they occurred.
TABLE 3-14. ERROR MESSAGE EXPLANATIONS
ERROR MESSAGE
EXPLANATION
General
Multiple Decimal places not supported Decimal key was pressed while to the right of the existing decimal point.
Exceeding Maximum Allowed Value
To check maximums, go to power-up screen, press
%
, highlight Max/Min Settings, press
!
.
Sign does not affect Protection Limits
Protection Limit field is positive or negative, numerical value does not need sign
Entry incomplete, Depress ENTER
key
When changing either the main channel or the protect channel, using the number keys, the
change must be applied to the output by pressing ENTER before moving on to change the
other channel.
Executing List
A waveform is currently executing via remote mode. Commands to change the output are
not accepted until list is stopped (send command VOLT:MODE FIXED).
Save Disabled, Request ignored
Save Display Change is disabled. From power-up screen, press
%
, highlight password
settings to enable.
EXTERNAL MODE enabled.
MODE key does not function while Mode is being controlled externally. Press
$
, high-
light External Mode and change to Disable.
F2 Key not valid
F2 only works when labeled. key was probably pressed inadvertently.
Password incorrect
Enter the correct password and press
!
. If password has been lost and can not be
recovered, refer to PAR. 3.6.3.12.1 to reset to factory defaults.
Setup Menu
F3 - Invalid
#
only works when labeled. Key was probably press inadvertently.
F4 - Invalid
$
only works when labeled. Key was probably press inadvertently.
Master/Slave Menu
Series or Parallel when not
STANDALONE
Mismatch between Unit type and Connection type. Automatically reverts to
Unit type: Standalone, Connection Type: standalone.
Calibration invalid - Multiple Unit
Consult Factory.
BITBUS option must be Disabled
Change CIIL mode to SCPI. From power-up screen press
%
highlight Interface Data For-
mat, press
!
, highlight SCPI, press
$
, then press
$
to save for power-up, then
cycle BOP off, then on.
Waveform Menu
External Reference Waveform not
accessible
Reference Input must be set to Internal for waveforms to be executed. Press
$
highlight
Reference Input, press
!
highlight Internal, press
$
, then
%
.
External Mode - Waveform not
accessible
External Mode must be disabled for waveforms to be executed. Press
$
highlight Exter-
nal Mode, press
!
highlight Disable, press
$
, then
%
.
External Limits - Waveform not
accessible
Protection Limit must be set to Internal for waveforms to be executed. Press
$
highlight
Protection Limit, press
!
highlight Internal, press
$
, then
%
.
Empty Entry - use F2 instead
Empty location cannot be modified using
!
Press
@
to start new waveform.
No Room to insert - delete a segment. No more than 10 segments are allowed in a waveform.
Unable to delete segment.
Try again. If still unable to delete segment, contact factory. for RMA.
Only EDIT allowed on Entry.
When viewing waveform, if name, count or protection levels are highlighted, press
!
to
modify the setting.
@
and
!
are not functional with these parameters.
MODE incorrect - Please Change
A voltage waveform was executed while the unit was in current mode or a current
waveform was executed while the unit was in voltage mode. Press MODE key to change to
correct mode, press STANDBY key to set output on then press
$
to run the waveform