beautypg.com

12 error message explanations, Table 3-14. error message explanations, Error message explanations -36 – KEPCO BOP 1KW-MG Operator Manual, Firmware Ver.4.12 and higher User Manual

Page 102: 14 error message explanations -36

background image

3-36

BOP HIPWR 031014

3.3.12

ERROR MESSAGE EXPLANATIONS

When a key press is not accepted, an error message is displayed at the bottom of the screen.
Although recovery from most operator errors is obvious and simple, Table 3-14 lists all the error
messages along with associated explanations as to why they occurred.

TABLE 3-14. ERROR MESSAGE EXPLANATIONS

ERROR MESSAGE

EXPLANATION

General

Multiple Decimal places not supported Decimal key was pressed while to the right of the existing decimal point.

Exceeding Maximum Allowed Value

To check maximums, go to power-up screen, press

%

, highlight Max/Min Settings, press

!

.

Sign does not affect Protection Limits

Protection Limit field is positive or negative, numerical value does not need sign

Entry incomplete, Depress ENTER
key

When changing either the main channel or the protect channel, using the number keys, the
change must be applied to the output by pressing ENTER before moving on to change the
other channel.

Executing List

A waveform is currently executing via remote mode. Commands to change the output are
not accepted until list is stopped (send command VOLT:MODE FIXED).

Save Disabled, Request ignored

Save Display Change is disabled. From power-up screen, press

%

, highlight password

settings to enable.

EXTERNAL MODE enabled.

MODE key does not function while Mode is being controlled externally. Press

$

, high-

light External Mode and change to Disable.

F2 Key not valid

F2 only works when labeled. key was probably pressed inadvertently.

Password incorrect

Enter the correct password and press

!

. If password has been lost and can not be

recovered, refer to PAR. 3.6.3.12.1 to reset to factory defaults.

Setup Menu

F3 - Invalid

#

only works when labeled. Key was probably press inadvertently.

F4 - Invalid

$

only works when labeled. Key was probably press inadvertently.

Master/Slave Menu

Series or Parallel when not
STANDALONE

Mismatch between Unit type and Connection type. Automatically reverts to
Unit type: Standalone, Connection Type: standalone.

Calibration invalid - Multiple Unit

Consult Factory.

BITBUS option must be Disabled

Change CIIL mode to SCPI. From power-up screen press

%

highlight Interface Data For-

mat, press

!

, highlight SCPI, press

$

, then press

$

to save for power-up, then

cycle BOP off, then on.

Waveform Menu

External Reference Waveform not
accessible

Reference Input must be set to Internal for waveforms to be executed. Press

$

highlight

Reference Input, press

!

highlight Internal, press

$

, then

%

.

External Mode - Waveform not
accessible

External Mode must be disabled for waveforms to be executed. Press

$

highlight Exter-

nal Mode, press

!

highlight Disable, press

$

, then

%

.

External Limits - Waveform not
accessible

Protection Limit must be set to Internal for waveforms to be executed. Press

$

highlight

Protection Limit, press

!

highlight Internal, press

$

, then

%

.

Empty Entry - use F2 instead

Empty location cannot be modified using

!

Press

@

to start new waveform.

No Room to insert - delete a segment. No more than 10 segments are allowed in a waveform.

Unable to delete segment.

Try again. If still unable to delete segment, contact factory. for RMA.

Only EDIT allowed on Entry.

When viewing waveform, if name, count or protection levels are highlighted, press

!

to

modify the setting.

@

and

!

are not functional with these parameters.

MODE incorrect - Please Change

A voltage waveform was executed while the unit was in current mode or a current
waveform was executed while the unit was in voltage mode. Press MODE key to change to
correct mode, press STANDBY key to set output on then press

$

to run the waveform