Procedure, Values are as follows, Follow these steps to set up an server ip address – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual
Page 70: Use the softkey as follows

Chapter 2
Configuration Settings
64
9650-0902-01 Rev. K
Values are as follows:
Procedure
Follow these steps to set up an server IP address.
Possible Values:
XXX.XXX.XXX.XXX where XXX is 0 to 255
Default:
000.000.000.000
Step
Action
1.
From the Select Configuration Categories screen,
select Wi-Fi, then press the Enter softkey.
The Wi-Fi Settings screen appears.
2.
Press the Next Item softkey until Server IP (Defib
History) appears and press the Change Value
softkey.
The Set Defib History Server IP Address screen
appears.
3.
Use the softkey as follows:
•
— moves the highlight up, down,
left, and right through the character selection
table.
•
Enter Char — enters the highlighted
character for each digit of the server IP.
•
Return — returns you to the Wi-Fi
Configuration screen. Press the Return
softkey when you are finished entering the
server IP.
Next
Item
Prev
Item
System Configuration
Wi-Fi Configuration
Defib History
Transfer Mode
DHCP/
Static IP
Both
Static IP
Main
Menu
Exit
Config
Category
List
Change
Value
Full Disclosure
Transfer Mode
Server
Page 1 of 7
System Configuration
Set Defib History Server IP Address
192.168.042.001
Return
Enter
Char
1
Cancel
4
7
2
5
8
3
6
9
0
<
<
.