Energy level: shock 2, Energy level: shock 3, Ped energy level shock 1 – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual
Page 31

Advisory Defib
9650-0902-01 Rev. K
25
Energy Level: Shock 2
This setting allows you to select the energy level used for the second shock delivered by the
defibrillator when the unit is:
•
in AED mode or Advisory mode and using pads
•
in Manual mode, using external paddles or pads, and Basic Auto Energy Escalation is set to
Yes
Values are as follows:
Energy Level: Shock 3
This setting allows you to select the energy level used for the third and all subsequent shocks
delivered by the defibrillator when the unit is:
•
in AED mode or Advisory mode and using pads
•
in Manual mode, using external paddles or pads, and Basic Auto Energy Escalation is set to
Yes
Values are as follows:
Ped Energy Level Shock 1
This setting allows you to select the energy level used for the first pediatric shock delivered by
the defibrillator when the unit is:
•
in Advisory mode and using OneStep™ Pediatric electrodes.
•
in Manual mode, using OneStep Pediatric electrodes
Ped Energy Level: Shock 1 is the energy level to which the unit defaults when you power on the
unit with pediatric electrode pads.
Values are as follows:
Warning!
The R Series will default to pediatric Shock 1, 2, and 3 energies only when ZOLL
®
OneStep pediatric electrodes are used. With pedi-padz
®
and pedi-padz II (or other non-
ZOLL pediatric electrodes), you are required to manually set the selected energy to
pediatric levels, per institutional protocols.
Possible Values:
Adult: 1-10, 15, 20, 30, 50, 75, 100, 120, 150, and 200 J
Default:
Adult: 150 J
Possible Values:
Adult: 1-10, 15, 20, 30, 50, 75, 100, 120, 150, and 200 J
Default:
Adult: 200 J
Possible Values:
Pediatric: 1-10, 15, 20, 30, 50, 70, 85, 100, 120, 150, and 200 J
Default:
Pediatric: 50 J