Display settings, Display time, Select trace 2 at power up – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual
Page 46

Chapter 2
Configuration Settings
40
9650-0902-01 Rev. K
Values are as follows:
Display Settings
The Display settings are attributes that control what and how things appear on the R Series
monitor and include:
•
Display Time
•
Select Trace 2 At Power Up
•
Select Trace 3 At Power Up
•
ECG Color
•
SpO2 Color
•
CPR Color
•
EtCO2 Color
•
NIBP Color
•
SPO2 at Power Up
Display Time
This setting specifies whether to display the elapsed time since the unit was turned on (Elapsed
Time), or the real time (Real Time), or not to display the time (Off).
The elapsed time continues to be counted for up to 10 seconds after you power off the unit. This
gives the operator adequate time to change the battery without resetting the elapsed timer.
Elapsed time is reset to zero (00:00) whenever the unit has been off for more than 10 seconds.
Values are as follows:
Select Trace 2 At Power Up
This setting allows you to set which waveform will appear in the Trace 2 position on the display
when you turn on the R Series unit.
Values are as follows:
Note:
When you set this setting to 3 Lead ECG, the Select Trace 3 At Power Up setting
automatically defaults to 3 Lead ECG.
Possible Values:
Yes, No
Default:
No
Possible Values:
Real Time, Off, Elapsed Time
Default:
Off
Possible Values:
Off, 3 Lead ECG, CPR, Filtered ECG, SpO2, EtCO2
Default:
Filtered ECG