Nibp auto interval default, Adult initial cuff inflation, Pediatric initial cuff inflation – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual
Page 50: Neonatal initial cuff inflation

Chapter 2
Configuration Settings
44
9650-0902-01 Rev. K
NIBP Auto Interval Default
This parameter allows you to set the default for the amount of time between automatic
measurements.
Values are as follows:
Adult Initial Cuff Inflation
This parameter allows you to set the default NIBP cuff inflation pressure (the pressure to which
the NIBP cuff inflates at the beginning of its first measurement after power up).
Values are as follows:
Pediatric Initial Cuff Inflation
This parameter allows you to set the default pediatric NIBP cuff inflation pressure (the pressure
to which the NIBP cuff inflates at the beginning of its first measurement after power up, if set to
pediatric).
Values are as follows:
Neonatal Initial Cuff Inflation
This parameter allows you to set the default NIBP cuff inflation pressure (the pressure to which
the NIBP cuff inflates at the beginning of its first measurement after power up, if set to
neonatal).
Values are as follows:
Possible Values:
2.5, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 45, 60, 90, 120 minutes
Default:
30 minutes
Possible Values:
mmHg: 120, 140, 160, 180, 200, 220, 240, 260
kPa: 16.0, 18.7, 21.3, 24.0, 26.7, 29.3, 32.0, 34.7
Default:
mmHg: 160
kPa: 21.3
Possible Values:
mmHg: 80, 90, 100, 110, 120, 130, 140, 150
kPa: 10.7, 12.0, 13.3, 14.7, 16.0, 17.3, 18.7, 20.0
Default:
mmHg: 120
kPa: 16.0
Possible Values:
mmHg: 60, 70, 80, 90, 100, 110, 120, 130
kPa: 8.0, 9.3, 10.7, 12.0, 13.3, 14.7, 16.0, 17.3
Default:
mmHg: 90
kPa: 12.0