Page 35

Advisory Defib
9650-0902-01 Rev. K
29
—
If the defibrillator disarms itself after it was charged (because the Defib Ready time
elapsed before a shock was delivered), the unit displays the PRESS ANALYZE prompt
until the ANALYZE softkey is pressed.
Note:
The Display DO CPR Message feature is operational only when the Auto Charge in
Advisory Mode and Auto Analyze options are both set to Yes.
Values are as follows:
CPR Message After No Shock Advised
This parameter determines which CPR message/voice prompt is issued during the CPR interval
after a No Shock Advised analysis result. The selected message is displayed throughout the
entire CPR interval.
Values are as follows:
CPR Message After Shock
This parameter determines which CPR message/voice prompt is issued during the CPR interval
after the delivery of the final shock in a stacked shock sequence. The selected message is
displayed throughout the entire CPR interval.
Values are as follows:
Possible Values:
Yes, No
Default:
Yes
Possible Values:
“If No Pulse, Perform CPR”, “Perform CPR”
Default:
Perform CPR
Possible Values:
“If No Pulse, Perform CPR”, “Perform CPR”
Default:
“Perform CPR”