Stacked Shocks
The setting specifies the number of stacked shocks that must be administered prior to the
activation of a CPR interval and works in conjunction with the Auto Analyze setting. If set to 3,
the unit will auto analyze 3 times before prompting the user. If set to 1, the unit will auto
analyze once.
Values are as follows:
Restart Analysis After CPR
When this parameter is set to Yes, the analysis starts automatically after the CPR interval
following each stacked shock sequence. Analyses do not automatically restart after a CPR
interval following No Shock Advised results. When this parameter is set to No, the PRESS
ANALYZE message is displayed, but the analysis does not restart.
Note:
This setting is always set to Yes in mode.
Values are as follows:
CPR Interval
This setting allows you to specify the duration of the CPR interval as described in the Display
DO CPR Message setting. Values are as follows:
Possible Values:
1, 3
Default:
1
Possible Values:
Yes, No
Default:
No
Possible Values:
1, 1.5, 2, 2.5, 3, 3.5, or 4 minutes
Default:
2 minutes