Restoring default values, Procedure – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual
Page 15

Restoring Default Values
9650-0902-01 Rev. K
9
Restoring Default Values
You can restore all configuration settings to their factory default settings. Only the display
language, if different from the default, does not change.
Procedure
Follow these steps to restore all configuration settings to default values:
If you want to review configuration settings, press the Change Config softkey. See “Accessing
and Changing Configuration Settings” on page 10 for more information. If you want to leave
System Configuration mode, press the Exit Config softkey. See “Leaving Configuration
Mode” on page 11 for more information.
Step
Action
1.
Power on the unit in System Configuration mode as
described in “Entering System Configuration Mode”
on page 2.
2.
Press the Restore Defaults softkey to reconfigure
settings to their default settings.
Once default settings are restored, the unit displays
the DEFAULTS RESTORED and PRESS CHANGE
CONFIG TO VIEW messages.
System Configuration
Change
Config
Exit
Config
Read
From
Card
Restore
Defaults
System Configuration
Defaults Restored
Press Change Config to View
Change
Config
Exit
Config
Read
From
Card
Restore
Defaults