Configuration Settings
24
9650-0902-01 Rev. K
Advisory Defib
The Advisory Defibrillation settings control functions related to performing defibrillation with
an R Series unit in Advisory mode and include:
•
Energy Level: Shock 1
•
Energy Level: Shock 2
•
Energy Level: Shock 3
•
Ped Energy Level Shock 1
•
Ped Energy Level Shock 2
•
Ped Energy Level Shock 3
•
Auto Analyze
•
Enable Check Pulse Prompt
•
Check Patient Prompt
•
Auto Charge in Advisory Mode
•
Enable Voice in Advisory Mode
•
Display DO CPR Message
•
CPR Message After No Shock Advised
•
CPR Message After Shock
•
Stacked Shocks
•
Restart Analysis After CPR
•
CPR Interval
Energy Level: Shock 1
This setting allows you to select the energy level used for the first shock delivered by the
defibrillator when the unit is:
•
in AED mode or Advisory mode and using pads
•
in Manual mode, using external paddles or pads
Energy Level: Shock 1 is the energy level to which the unit defaults when you power on the
unit with external paddles or pads connected.
Values are as follows:
Possible Values:
Adult: 1-10, 15, 20, 30, 50, 75, 100, 120, 150, and 200 J
Default:
Adult: 120 J