Alarm limits at power up – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual
Page 41

Alarms Settings
9650-0902-01 Rev. K
35
Alarm Limits at Power Up
This option allows you to enable or disable, and set high and low threshold values for the ECG
Heart Rate, and the SpO
2
alarms on the R Series unit. All alarms listed below (Heart Rate and
optional devices SpO
2
, EtCO
2
, NIBP) are enabled by default. Values are as follows:
Possible Values:
Heart Rate:
High: 60 - 280 beats per minute (BPM) inclusive in 5 BPM increments
Low: 20 - 100 beats per minute (BPM) inclusive in 5 BPM increments
Optional Devices:
SpO
2
:
High: 72-100% in 1% increments
Low: 70-98% in 1% increments
EtCO
2
:
High: 5-100 mmHg in increments of 5
0.6-13.3 kPa in increments of .67
0.6-13.1% in increments of .66
Low: 0-95 mmHg in increments of 5
0.0-12.6 kPa in increments of .67
0-12.5% in increments of 0.66
Resp Rate:
High: 5-150 breaths per minute in increments of 5
Low: 0-100 breaths per minute in increments of 5
NIBP Sys:
Adult:
High: 80-260 mmHg in increments of 5
10.7-34.7 kPa in increments of 0.67
Low: 40-140 mmHg in increments of 5
5.3-18.7 kPa in increments of 0.67
Pediatric:
High: 80-160 mmHg in increments of 5
10.7-21.3 kPa in increments of 0.67
Low: 35-140 mmHg in increments of 5
4.7-18.7 kPa in increments of 0.67
Neonatal:
High: 60-130 mmHg in increments of 5
8.0-17.3 kPa in increments of 0.67
Low: 25-120 mmHg in increments of 5
3.3-16.0 kPa in increments of 0.67