Enable readiness test audible alert (future use), One step padz installed – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual
Page 56

Chapter 2
Configuration Settings
50
9650-0902-01 Rev. K
Enable Readiness Test Audible Alert (Future Use)
This setting allows you to specify whether the unit emits a beep when the automatic device
check fails.
Values are as follows:
One Step Padz Installed
This setting allows you to specify whether the R Series unit tests the R Series OneStep
electrodes as part of the readiness test check, and which electrodes it tests.
•
If this setting is set to any of its values (including No or Base), and CPR, MwP or Complete
electrodes are detected, they will be tested and the resulting status (pass/fail) is included in
the readiness result.
•
If this setting is set to CPR/MwP/Complete, and none of these OneStep electrodes are
connected, or if the cable is not connected to the unit, the readiness result will indicate a
failure (Electrode: Not Connected).
•
If this setting is set to Base or No, and the cable is correctly connected to a short (Base
electrode package, test port, or paddles in the wells), the unit will perform the readiness test
and not fail based on electrode type.
Values are as follows:
Note:
The OneStep Padz Installed option should be set to Base unless you are using OneStep
electrodes marked Rev. B or above:
8900-0212-XX Rev. B or above for OneStep Pacing
8900-0213-XX Rev. B or above for OneStep CPR
8900-0214-XX Rev. B or above for OneStep Complete
Possible Values:
No
Default:
No
Possible Values:
CPR/MwP/Complete, No, Base
Default:
CPR/MwP/Complete