Automation Command and Query Reference Manual - Control Reference

AcquiredSegments

Integer

From 0 to 100000 step 1

Range

AcquisitionDuration

Double

From 1e-012 to 1e+012 step 1e-015

Range

Queries the duration of the last completed acquisition. The result may depend on the spacing of the
triggers in sequence mode, and it may depend on the number of averages when a channel is
in averaging mode.

Description

' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")

' Obtain the duration of the last completed acquistion.
AcqDuration = app.Acquisition.Horizontal.AcquisitionDuration
MsgBox AcqDuration

Example

ExtClockFrequency

Double

From 0.001 to 100000 step 0.001

Range

ExtCoupling

Enum

Specific to WR and WS series of the scope.
Sets/Queries coupling of external trigger/clock input.

Description

Values

DC1M

DC50

HorOffset

Double

From -1e-006 to 2e-006 step 4e-009

Range

Sets/Queries the horizontal position of the trigger time, relative to the origin set by HorOffsetOrigin, in
seconds. Positive to the right, negative to the left. The setting resolution is about 1% to 2&.

Description

' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")

' Set the horizontal trigger offset to 200 ns.
app.Acquisition.Horizontal.HorOffset = 2.0e-7

Example