Operator’s manual – Teledyne LeCroy WaveAce 1000_2000 User Manual
Page 99

Operator’s Manual
922131-00 Rev A
93
Electromagnetic Emissions:
CISPR 11:2003, Radiated and Conducted Emissions Group 1, Class A
2 3
EN 61000-3-2:2006 Harmonic Current Emissions, Class A
EN 61000-3-3/A2:2005 Voltage Fluctuations and Flickers, Pst = 1
Electromagnetic Immunity:
EN 61000-4-2:2001 Electrostatic Discharge, 4 kV contact, 8 kV air, 4 kV
vertical/horizontal coupling planes
4
EN 61000-4-3:2006 RF Radiated Electromagnetic Field, 3 V/m, 80-1000 MHz; 3
V/m, 1400 MHz - 2 GHz; 1 V/m, 2 GHz - 2.7 GHz
4
EN 61000-4-4:2004 Electrical Fast Transient/Burst, 1 kV on power supply lines,
0.5 kV on I/O signal data and control lines
4
EN 61000-4-5:2006 Power line Surge, 1 kV AC Mains, L-N, L-PE, N-PE
4
EN 61000-4-6:2007 RF Conducted Electromagnetic Field, 3 Vrms, 0.15 MHz -
80 MHz
4
EN 61000-4-11:2004 Mains Dips and Interruptions, 0%/1 cycle, 70%/25 cycles,
0%/250 cycles
4 5
1 To ensure compliance with all applicable EMC standards, high quality shielded
interface cables should be used.
2 Emissions which exceed the levels required by this standard may occur when the
oscilloscope is connected to a test object.
3 This product is intended for use in nonresidential areas only. Use in residential areas
may cause electromagnetic interference.
4 Meets Performance Criteria “B” limits of the respective standard: during the
disturbance, product undergoes a temporary degradation or loss of function or
performance which is self-recoverable.
5 Performance Criteria “C” applied for 70%/25 cycle voltage dips and for 0%/250 cycle
voltage interruption test levels per EN61000-4-11.
European Contact:
Teledyne LeCroy Europe GmbH
Waldhofer Str 104
D-69123 Heidelberg
Germany
Tel: (49) 6221 82700