Appendix I: Data Output Formats

6/12

58

51214 Rev. 7


Gross Weight – Motion – decimal points included




Net Weight – No Motion – decimal points included




Net Weight – Motion – decimal points included




PLC

Instrument Setup Protocol: 19,200 Baud, 8 Character bits, 1 Stop bit, and no parity.
This is a continuous output that transmits data on every other A/D update
(approximately 32
transmissions per second.

The output string is 12 characters, seen as "wwwwwwwuumm" where,

= Weight including decimal point and negative sign.

Negative sign precedes most significant digit and may be preceded by space

characters.

= units "lb", "kg", "oz", or "g ". "oz" may indicate ounces, or lb/ounces.

= mode "GR" or "NT" for stable weight, or "gr" or "nt" for motion.

= end of text.


Example 1: Gross - lbs - Weight Stable

Example 2: Gross - lbs - Weight in Motion

Example 3: Net - kgs - Weight Stable

Example 4: Net - kgs - Weight in Motion


Note: Once PLC option is selected and protocol has been programmed, the installing
technician's setup is complete. Confirmation of output can be performed by
monitoring the COM port with a terminal program such as Windows Hyper Terminal.

Note: Interface of the HMS Communications device, power supply to same device,
and programming of the PLC is strictly the responsibility of the customer.