Section 2: Front Panel programming

6/12

14

51214 Rev. 7


CLr.ACC: Clearing the Accumulator:
Conditional upon programming the operating mode to accumulation, peak hold, piece
count, or piece count & total. The display will indicate CLr.ACC. Press the ZERO key,
then the UNITS key to toggle the available settings. Press the ZERO key to enter the
selection. Available selections are:

Clr.YES (Clear the Accumulator)

CLr.NO (Do NOT Clear the Accumulator)

Note:
Pressing ZERO will clear the Accumulator and return to the Net/Gross mode.
Pressing UNITS will NOT clear the Accumulator and return to the Net/Gross mode.

Press the Zero key for 3 seconds

Conditional upon programming the operating mode to accumulation.
The FB2250 will display total number of pieces and weight, then time out (30 sec).

Conditional upon programming the operating mode to piece count, or piece count &
total. The FB2250 will display total number of pieces, then time out (30 sec).

Conditional upon programming the operating mode to peak hold stable or peak hold
unstable. The FB2250 will display time, date, peak weight, and time out (30 sec).

Note: Pressing the UNITS key will skip time out and immediately exit to the weigh
mode.

4.20: Setting the 4-20 mA.

Press the Units key for 3 seconds

Conditional upon enabling 4-20 mA.

Gross / Net:
Programming the weigh mode the output will track: Select Gross or Net.
The 4-20 mA analog output will track the selection. It will continue to track the
selection regardless of the currently selected weigh mode of the FB2250.
Press the ZERO key, then the UNITS key to toggle the available settings. Press the
ZERO key to enter the selection. Available selections are Gross or Net.

4.20 Lo: Programming the weight value associated with 4 mA: The display will
indicate 4.20 Lo: Press the ZERO key. The display will indicate the zero weight
“XXXXXX”. Press the UNITS key to increment the blinking digit, press the ZERO key
to advance to the next digit, repeat until entry is complete. After entering the least
significant digit, press the ZERO key. This weight entry will result in a 4 mA output.

4.20 Hi: Programming the weight value associated with 20 mA: The display will
indicate 4.20 Hi: Press the ZERO key. The display will indicate the maximum weight
“XXXXXX”. Press the UNITS key to increment the blinking digit, press the ZERO key
to advance to the next digit, repeat until entry is complete. After entering the least
significant digit, press the ZERO key. This weight entry will result in a 20 mA output.