Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide
41
Using IO Interfaces
Using LAN
Using Telnet On a PC With a Host/Port Setting Menu GUI
1.
On your PC, click
Start > Run.
2.
Type telnet then click the
OK button. The Telnet connection screen will be displayed.
3.
Click on the
Connect menu then select Remote System. A connection form is displayed. See
4.
Enter the
hostname, port number, and TermType then click Connect.
•
Host Name
−signal generator’s IP address or hostname
•
Port
−5023
•
Term Type
−vt100
5.
At the SCPI> prompt, enter SCPI commands. Refer to
6.
To signal device clear, press
Ctrl-C.
7.
Select
Exit from the Connect menu to end the Telnet session.
Figure 2-6
Telnet Window (Windows 2000)
Using Telnet On Windows 2000
1.
On your PC, click
Start > Run.
2.
Type telnet in the run text box, then click the
OK button. The Telnet connection screen will be
displayed. See
(Windows 2000).
3.
Type open at the prompt and then press the
Enter key. The prompt will change to (to).
4.
At the (to) prompt, enter the signal generator’s IP address followed by a space and 5023, which
is the Telnet port associated with the signal generator.
5.
At the SCPI> prompt, enter SCPI commands. Refer to commands shown in
.
6.
To escape from the SCPI> session type Ctrl-].
7.
Type quit at the prompt to end the Telnet session.