Baseband operation status group, P (see, Baseband operation status – Agilent Technologies N5183A MXG User Manual
Page 172
162
Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide
Programming the Status Register System
Status Groups
Baseband Operation Status Group
NOTE
This status group does not apply to the E4428C, E8257D, and the E8663B, and if queried,
returns zero. See
for more information.
This status group does not apply to the N5181A/82A/83A. (If queried, the signal generator
will not respond.)
The Baseband Operation Status Group is used to determine the specific event that set bit 10 in the
Standard Operation Status Group
. This group consists of the
Baseband Operation Condition Register
,
the
Baseband Operation Transition Filters (negative and positive)
, and the
Baseband Operation Event Enable Register
This is the named
status register for
the E4438C.
However, not all
signal generator
models use all of
the shown events
(i.e. some use only
a subset of the
E4438C’s status
registers).