Data Requirements
•
The signal generator requires a file with a minimum of 60 bytes
•
For every data bit (bit 0), you must provide 7 bits of control information (bits 1- 7).
ERROR - 223, Too much data
PRAM download exceeds the size of PRAM memory.
Either use a smaller pattern or get more memory by ordering the appropriate
hardware option.
Table 8-14 PRAM Data Byte
Bit
Function
Value
Comments
0
Data
0/1
This is the data (payload) bit. It is “unspecified” when burst (bit 2) is set to 0.
1
Reserved
0
Always 0
2
Burst
0/1
1 = RF on
0 = RF off
For non- bursted, non- TDMA systems, to have a continuous signal, set this bit to 1 for all
bytes. For framed data, set this bit to 1 for on timeslots and 0 for off timeslots.
3
Reserved
0
Always 0
4
Reserved
1
Always 1
5
Reserved
0
Always 0
6
EVENT1
Output
0/1
To have the signal generator output a single pulse at the EVENT 1 connector, set this bit
to 1. Use this output for functions such as a triggering external hardware to indicate when
the data pattern begins and restarts, or creating a data- synchronous pulse train by
toggling this bit in alternate bytes.
7
Pattern Reset
0/1
0 = continue to next sequential memory address.
1 = end of memory and restart memory playback.
This bit is set to 0 for all bytes except the last byte of PRAM. To restart the pattern, set
the last byte of PRAM to 1.
Table 8-13 PRAM Download - Symptoms and Causes
Symptom