Agilent Technologies N5183A MXG User Manual
Page 163
![background image](/manuals/34672/163/background.png)
Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide
153
Programming the Status Register System
Accessing Status Register Information
Table 4-2
Effects of :STATus:PRESet
Register
a
a.Table reflects :STAT:PRES values for an E4438C with options 001, 002, 601, or 602 and UN7. To determine the registers that apply to your
signal generator, refer to
and
.
Value after
:STATus:PRESet
:STATus:OPERation:ENABle
0
:STATus:OPERation:NTRansition
0
:STATus:OPERation:PTRransition
32767
:STATus:OPERation:BASeband:ENABle
0
:STATus:OPERation:BASeband:NTRansition
0
:STATus:OPERation:BASeband:PTRransition
32767
:STATus:QUEStionable:CALibration:ENABle
32767
:STATus:QUEStionable:CALibration:NTRansition
32767
:STATus:QUEStionable:CALibration:PTRansition
32767
:STATus:QUEStionable:ENABle
0
:STATus:QUEStionable:NTRansition
0
:STATus:QUEStionable:PTRansition
32767
:STATus:QUEStionable:FREQuency:ENABle
32767
:STATus:QUEStionable:FREQuency:NTRansition
32767
:STATus:QUEStionable:FREQuency:PTRansition
32767
:STATus:QUEStionable:MODulation:ENABle
32767
:STATus:QUEStionable:MODulation:NTRansition
32767
:STATus:QUEStionable:MODulation:PTRansition
32767
:STATus:QUEStionable:POWer:ENABle
32767
:STATus:QUEStionable:POWer:NTRansition
32767
:STATus:QUEStionable:POWer:PTRansition
32767
:STATus:QUEStionable:BERT:ENABle
32767
:STATus:QUEStionable:BERT:NTRansition
32767
:STATus:QUEStionable:BERT:PTRansition
32767