beautypg.com

Data questionable modulation event enable, Register – Agilent Technologies N5183A MXG User Manual

Page 186

background image

176

Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide

Programming the Status Register System
Status Groups

Data Questionable Modulation Event Enable Register

The Data Questionable Modulation Event Enable Register lets you choose which bits in the Data
Questionable Modulation Event Register set the summary bit (bit 7 of the Data Questionable
Condition Register) to 1.

Command:

STATus:QUEStionable:MODulation:ENABle

where is the sum of the decimal values

of the bits you want to enable.

Example:

Enable bit 4 and bit 3 so that whenever either of those bits are set to 1, the Data Questionable
Modulation summary bit of the Data Questionable Condition Register is set to 1. Send the command
STAT:QUES:MOD:ENAB 520

(16 + 8).

Query:

STATus:QUEStionable:MODulation:ENABle?

Response:

Decimal value of the sum of the bits previously enabled with the
STATus:QUEStionable:MODulation:ENABle

command.

This manual is related to the following products: