5 Maintenance

Models 3010P

5-4: Part II

2. With span gas flowing, read the raw output of the cell from the

System

function display.

3. Divide the raw output reading by the percent oxygen

concentration of your span gas.

If the quotient is less than the Index value for the cell class you are

using, replace the cell.

Table 5-1: Cell Indices

Cell Class

Index

A-3

1.818

A-5

1.818

B-1

4.545

B-3

3.716

B-5

1.244

B-7

1.515

C-3

2.488

C-5

0.606

5.2.3 Removing the Micro-Fuel Cell

WARNING: DO NOT TOUCH THE SENSING SURFACE OF THE

CELL. IT IS COVERED WITH A DELICATE TEFLON
MEMBRANE THAT CAN LEAK CAUSTIC AND COR-
ROSIVE CHEMICALS WHEN PUNCTURED.

The Micro-Fuel cell is located inside the housing in a nylon cell

block. (Some models may have a stainless steel block). See Figure 5-2.

To remove an existing cell:

1. Remove power to the instrument by unplugging the power cord

at the power source.

2. Rotate the housing cover counterclockwise until it is free from

the housing, and then remove it.

3. Pull up on the Probe, with a slight rocking motion, to release it

from the Probe Receptacle.

4. Do Not remove the O-rings unless they are worn and no longer

hold the Probe tightly. (If worn, replace them.)