Teledyne 3010PA - Split architecture percent oxygen analyzer User Manual
Page 40
Oxygen Analyzer
Part I: Control Unit
Part I: 4-9
4.3.5 System Self-Diagnostic Test
The Model 3010PA has a built-in self-diagnostic testing routine. Pre-
programmed signals are sent through the power supply, output board and
sensor circuit. The return signal is analyzed, and at the end of the test the
status of each function is displayed on the screen, either as
OK
or as a
number between 1 and 3. (See System Self Diagnostic Test in chapter 5 for
number code.)
Note: Remote Probe connector must be connected to the Analysis
Unit, or sensor circuit will not be properly checked.
The self diagnostics are run automatically by the analyzer whenever the
instrument is turned on, but the test can also be run by the operator at will.
To initiate a self diagnostic test during operation:
Press the
System
button to start the
System
function.
Contrast AutoCal
PSWD Logout More
Use the < > arrow keys to blink
More
, then press
Enter
.
Version SelfTest
Use the < > arrow keys again to move the blinking to the
Self–Test
function. The screen will follow the running of the diagnostic.
RUNNING DIAGNOSTIC
Testing Preamp 83
During preamp testing there is a countdown in the lower right corner of
the screen. When the testing is complete, the results are displayed.
Power: OK Analog: OK
Preamp: 3
The module is functioning properly if it is followed by
OK
. A number
indicates a problem in a specific area of the instrument. Refer to chapter 5
Maintenance for number-code information. The results screen alternates for a
time with:
Press Any Key
To Continue...
Then the analyzer returns to the initial System screen.
Contrast Function is
DISABLED
(Refer to Section 1.3)