Part II: Analysis Unit

Part II: 4-3

Table 4-1: Self Test Failure Codes

Power

0

OK

1

5 V Failure

2

15 V Failure

3

Both Failed

Analog

0

OK

1

DAC A (0–1 V Concentration)

2

DAC B (0–1 V Range ID)

3

Both Failed

Preamp

0

OK

1

Zero too high

2

Amplifier output doesn't match test input

3

Both Failed

The results screen alternates for a time with:

Press Any Key

To Continue...

Then the analyzer returns to the initial System screen.

4.5

Cell Failure Checks

Cell failure is covered in detail in Part I: Control Units, section

4.4.1.3, Cell Failure. Cell replacement is covered Part II: Analysis Units
chapter 5, Maintenance.

When the sensor in the 3010P begins to fail, the analyzer will usually

require more and more frequent calibration. If recalibration raises the
readings temporarily only, suspect the cell.

You can check the output of the cell itself by scrolling the MAIN

MENU TO

SENSOR.

When you ENTER the function, the sensor report screen appears.

RAW CELL OUTPUT

###

µµµµµ

A