4 Operation

Model 3010P

4-4: Part II

The “good” reading depends on the class of cell your analyzer is

using. Although the B-1 cell is standard in the 3010P, check Specific
Model Information
in the Front Matter in this manual for the class of
cell you purchased
. Then check Cell Replacement in chapter 5 Mainte-
nance
, and do the prescribed calculations. If a weak cell is indicated,
replace the cell as described in chapter 5.

After correcting the condition, reset the Cell Fail Alarm by taking the

analyzer into, and then back out of, STANDBY.