beautypg.com

Maintenance – Teledyne 3010PA - Split architecture percent oxygen analyzer User Manual

Page 50

background image

Part I: Control Unit

Maintenance 5

Part I: 5-1

Maintenance

Aside from normal cleaning and checking for leaks at the gas connec-

tions, routine maintenance is limited to replacing Micro-Fuel cells and
fuses, and recalibration.

Checking for leaks, replacing Micro-Fuel cells, and replacing fuses in

the Analysis Unit are covered in Part II, Chapter 5. For recalibration, see
Part I, section 4.4 Calibration.

WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS

MANUAL.

5.1

Fuse Replacement

1. Place small screwdriver in notch, and pry cover off, as shown in

Figure 5-1.

Figure 5-1: Removing Fuse Block from Housing

2. To change between American and European fuses, remove the

single retaining screw, flip Fuse Block over 180 degrees, and
replace screw.