v

Oxygen Analyzer

Oxygen Analyzer

Oxygen Analyzer

Oxygen Analyzer

Oxygen Analyzer

Preface

Overview

The Analytical Instruments Model 3010PA Percent Oxygen Analyzer is

a versatile microprocessor-based instrument for detecting oxygen in a
variety of background gases. It is a “split architecture” instrument. This
means that a general purpose Control Unit, designed for nonhazardous areas
only, remotely controls a specially designed Analysis Unit, or remote probe,
that can operate in a hazardous area.

Part I of this manual covers the Model 3010PA General Purpose flush-

panel and/or rack-mount Control Unit only. This Control Unit is for indoor
use in a nonhazardous environment. The Analysis Units (or Remote Probes)
they control, can be designed for a variety of hazardous environments. Part II
of this manual covers the 3010P Analysis Unit.

Typical Applications

A few typical applications of the Model 3010PA are:

Monitoring inert gas blanketing

Air separation and liquefaction

Chemical reaction monitoring

Semiconductor manufacturing

Petrochemical process control

Quality assurance

Gas analysis certification.

Model and Part Number Designations

The part numbers are the most specific identification. When using this

manual for operation, maintenance, or ordering parts, check the part numbers