beautypg.com

Teledyne 3010PA - Split architecture percent oxygen analyzer User Manual

Page 65

background image

2-4: Part II

2 Operational Theory

Model 3010

P

(These reactions will hold as long as no gaseous components capable of

oxidizing lead—such as iodine, bromine, chlorine and fluorine—are present
in the sample.)

The output of the fuel cell is limited by (1) the amount of oxygen in the

cell at the time and (2) the amount of stored anode material.

In the absence of oxygen, no current is generated.

2.2.4 The Effect of Pressure

In order to state the amount of oxygen present in the sample in parts-

per-million or a percentage of the gas mixture, it is necessary that the sample
diffuse into the cell under constant pressure.

If the total pressure increases, the rate that oxygen reaches the cathode

through the diffusing membrane will also increase. The electron transfer, and
therefore the external current, will increase, even though the oxygen concen-
tration of the sample has not changed. It is therefore important that the
sample pressure at the fuel cell (usually vent pressure) remain relatively
constant between calibrations.

2.2.5 Calibration Characteristics

Given that the total pressure of the sample gas on the surface of the

Micro-Fuel Cell input is constant, a convenient characteristic of the cell is
that the current produced in an external circuit is directly proportional to the
rate at which oxygen molecules reach the cathode, and this rate is directly
proportional to the concentration of oxygen in the gaseous mixture. In other
words it has a linear characteristic curve, as shown in Figure 2-3. Measuring
circuits do not have to compensate for nonlinearities.

In addition, since there is zero output in the absence of oxygen, the

characteristic curve has close to an absolute zero. Therefore, the cell itself
does not need to be zeroed. (The electronic circuits are zeroed automatically
when the instrument power is turned on.)