Part I: 2-3

Oxygen Analyzer

Part I: Control Unit

The digital concentration signal—along with input from the control

panel—is processed by the microprocessor, and appropriate control signals
are directed to the display, alarms and communications port as well as to
the optional gas control valves in the Analysis Unit.

The same digital information is also sent to a 12 bit digital to analog

converter that produces the 0-1 V dc and the optional 4-20 mA dc analog
concentration signal outputs, and the analog range ID outputs.

The microprocessor monitors the power supply, and activates the

system failure alarm if a malfunction is detected.

2.3 Temperature Control

For accurate analysis this instrument is temperature controlled not to

fall beneath a certain temperature. This temperature is 22

o

F. This is to

prevent the sensor from freezing in cold environments.