Teledyne 3010PA - Split architecture percent oxygen analyzer User Manual

Page 100

background image

Appendix: A-11

Oxygen Analyzers

Appendix

Zero: Settling: Man

To Begin

Press

Enter

to begin the zero calibration. After a few seconds the first

of five zeroing screens appears. The number in the upper left hand corner is
the first-stage zero offset. The microprocessor samples the output at a prede-
termined rate. It calculates the differences between successive samplings and
displays the rate of change as

Slope=

a value in parts per million per second

(ppm/s).

#### % Zero

Slope=#### ppm/s

NOTE: It takes several seconds for the true Slope value to display. Wait about

10 seconds. Then, wait until Slope is sufficiently close to zero before
pressing

Enter to finish zeroing. Slope is given in ppm/s.

Generally, you have a good zero when

Slope

is less than 0.05 ppm/s

for about 30 seconds. When

Slope

is close enough to zero, press

Enter

. In a

few seconds, the screen will update.

Once span settling completes, the information is stored in the micropro-

cessor, and the instrument automatically returns to

Analyze

mode.