4 Operation

Model 3010

PA

4-16: Part I

Use the

∆∇

arrow keys to enter the upper value of the range (all ranges

begin at 0 %). Repeat for each range you want to set. Press

Enter

to accept

the values and return to

Analyze

mode. (See note below.)

Note: The ranges must be increasing from low to high, for example,

if range 1 is set for 0–10 % and range 2 is set for 0–100 %,
range 3 cannot be set for 0–50 % since it is lower than range 2.

4.6.2 Fixed Range Analysis

The autoranging mode of the instrument can be overridden, forcing the

analyzer DC outputs to stay in a single predetermined range.

To switch from autoranging to fixed range analysis, enter the range

function by pressing the

Range

button on the front panel.

Use the

< > arrow keys to move the blinking over

AUTO

.

Use the

∆∇

arrow keys to switch from

AUTO

to

FX/LO

,

FX/MED,

or

FX/HI

to set the instrument on the desired fixed range (low, medium, or

high).

L—### M—####

H—##### Mode—FX/LO

or

L—### M—####

H—##### Mode—FX/MED

or

L—### M—####

H—##### Mode—FX/HI

Press

Escape

to re-enter the

Analyze

mode using the fixed range.

NOTE:When performing analysis on a fixed range, if the oxygen

concentration rises above the upper limit (or default value) as
established by the operator for that particular range, the
output saturates at 1 V dc (or 20 mA). However, the digital
readout and the RS-232 output continue to read the true value
of the oxygen concentration regardless of the analog output
range.