3 Installation

Model 3010

P

3-6: Part II

Figure 3-4: Control Unit (CU) to Analysis Unit (AU) Connector Cable

If you use your own gas control valves, use the interconnect diagram in

Figure 3-5. (See drawing D-64950 for wire recommendations.)

Figure 3-5: Remote Probe Connector Pinouts

The voltage from the solenoid outputs is nominally 0 V for the OFF

and 15 V dc for the ON conditions. The maximum combined current that
can be pulled from these output lines is 100 mA. (If two lines are ON at the
same time, each must be limited to 50 mA, etc.)

If more current and/or a different voltage is required, use a relay, power

amplifier, or other matching circuitry to provide the actual driving current.
Note that each individual line has a series FET with a nominal ON resistance
of 5 ohms (9 ohms worst case). This can limit the obtainable voltage, de-
pending on the load impedance applied. See Figure 3-6.