Part I: Control Unit
Part I: 4-11
Version SelfTest
Cell Output: ### µA
The “good” reading depends on the class of cell your analyzer is using.
Although the B-1 cell is standard in the 3010PA, check Specific Model
Information in the Front Matter in this manual for the class of cell you
purchased.
Then check Cell Replacement in Part II Analysis Units, chapter 5
Maintenance, and do the prescribed calculations. If a weak cell is indicated,
replace the cell as described there in chapter 5.
4.4.2 Span Cal
The
Span
button on the front panel is used to span calibrate the ana-
lyzer. Span calibration can be performed using the automatic mode, where
an internal algorithm compares consecutive readings from the sensor to
determine when the output matches the span gas concentration. Span cali-
bration can also be performed in manual mode, where the operator deter-
mines when the span concentration reading is acceptable and manually exits
the function.
4.4.2.1
Auto Mode Spanning
Press
Span
to enter the span function. The screen that appears allows
you to select whether the span calibration is to be performed automatically or
manually. Use the
∆∆∆∆∆∇
arrow keys to toggle between
AUTO
and
MAN
span
settling. Stop when
AUTO
appears, blinking, on the display.
Span: Settling: AUTO
Press
Enter
to move to the next screen.
Span Val: 20.90
Use the
∆∆∆∆∆∇
arrow keys to enter the oxygen-concentration mode. Use
the < > arrow keys to blink the digit you are going to modify. Use the
∆∆∆∆∆∇
arrow keys again to change the value of the selected digit. When you have
finished typing in the concentration of the span gas you are using (20.90 if
you are using air), press
Enter
to begin the Span calibration.
#### % Span
Slope=#### ppm/s
The beginning span value is shown in the upper left corner of the
display. As the span reading settles, the screen displays and updates informa-