background image

Oxygen Analyzer

Part I: Control Unit

Part I: 4-1

Operation

4.1

Introduction

Once the analyzer has been installed, configure it for your process. To

do this you can:

Set system parameters—

Specify a password, if desired, requiring operator to log in.

Establish and start an automatic calibration cycle, if desired.

Calibrate the instrument.

Define the three user selectable analysis ranges. Then choose
autoranging or select a fixed range of analysis, as required.

Set alarm setpoints, and modes (latching, failsafe, etc).

Before configuration these default values are in effect:

PARAMETER

DEFAULT

LO Range

1 %

MED Range

5 %

HI Range

10 %

Auto Ranging

ON

Alarm Relays

10 %
(Defeated, HI, Not failsafe, Not latching)

Span

20.9 %
(Auto, every 0 days at 0 hours)

Zero

(Auto, every 0 days at 0 hours).

If you choose not to use password protection, the default password is

automatically displayed on the password screen when you start up, and you
simply press

Enter

for access to all functions of the analyzer.