Oxygen Analyzer

Part I: Control Unit

Part I: 4-5

4.3.2 Setting up an Auto-Cal

When the proper calibration gases are connected (see chapter 3, instal-

lation), the Analyzer can cycle itself through a sequence of steps that auto-
matically zero and span the instrument.

Note: If you require highly accurate Auto-Cal timing, use external

Auto-Cal control where possible. The internal clock in the
Model 3010PA is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.

To setup an

Auto–Cal

cycle:

Choose

System

from the Function buttons. The VFD will display five

subfunctions.

Contrast Auto—Cal

PSWD Logout More

Use < > arrows to blink

Auto—Cal

, and press

Enter

. A new screen for

Span/Zero

set appears.

Span OFF Nxt: 0d 0h

Zero OFF Nxt: 0d 0h

Press < > arrows to blink

Span

(or

Zero

), then press

Enter

again. (You

won’t be able to set

OFF

to

ON

if a zero interval is entered.) A

Span

Every

... (or

Zero Every

...) screen appears.

Span Every 0 d

Start 0 h from now

Use

∆∆∆∆∆∇

arrows to set an interval value, then use < > arrows to move to

the start-time value. Use

∆∆∆∆∆∇

arrows to set a start-time value.

To turn

ON

the

Span

and/or

Zero

cycles (to activate

Auto-Cal

): Press

System

again, choose

Auto—Cal,

and press

Enter

again. When the

Span/

Zero

values screen appears, use the < > arrows to blink the

Span

(or

Zero

)

OFF

/

ON field. Use

∆∆∆∆∆∇

arrows to set the

OFF

/

ON field to ON. You can

now turn these fields

ON because there is a nonzero span interval defined.

4.3.3 Password Protection

If a password is assigned, then setting the following system parameters

can be done only after the password is entered: span and zero settings,
alarm setpoints, analysis range definitions, switching between autoranging
and manual override, setting up an auto-cal, and assigning a new password.
However, the instrument can still be used for analysis or for initiating a self-
test without entering the password.

Contrast Function is

DISABLED

(Refer to Section 1.3)