Teledyne 3010PA - Split architecture percent oxygen analyzer User Manual
Page 72
Oxygen Analyzer
Part II: Analysis Unit
Part II: 3-3
For pressurized service, use the restrictor without the blue dot and union
from the restrictor kit and attach it to the Sample In port. The small circular
orifice should face away from the back of the unit (against the direction of
gas flow). Use the restrictor without the blue dot sticker in the same manner
for low pressure applications (less than 5 psig).
For vacuum service (5-10 in Hg), use the restrictor without the blue dot
sticker and union but attach it to the Exhaust Out port. The small circular
orifice should face toward the back of the unit (against the direction of gas
flow).
Remove the blue sticker from the restrictor before using.
WARNING: Operating the unit without restrictors can cause
damage to the micro-fuel cell.
Figure 3-2 shows the Model 3010P Gas Connector panel. The inlets for
zero and span gas are included only with the –C option.
Figure 3-2: Gas Connector Panel of the Model 3010P
The unit is manufactured with
1
/
4
inch tube fittings, and 6 mm adapters
are supplied for metric system installations. For a safe connection:
1. Insert the tube into the tube fitting, and finger-tighten the nut until
the tubing cannot be rotated freely, by hand, in the fitting. (This
may require an additional
1
/
8
turn beyond finger-tight.)
2. Hold the fitting body steady with a backup wrench, and with
another wrench rotate the nut another 1
1
/
4
turns.