beautypg.com

Agilent Technologies E8267D PSG User Manual

Page 129

background image

Chapter 3

115

Basic Digital Operation

Triggering Waveforms

3.

Configure the carrier signal output:

Set the desired frequency.

Set the desired amplitude.

Turn on the RF output.

4.

Select a waveform for playback (sequence or segment):

a.

Preset the signal generator.

b.

Press

Mode

>

Dual ARB

>

Select Waveform

.

c.

Highlight a waveform file (for this example, SINE_TEST_WFM).

d. Press

Select Waveform

.

5.

Select the waveform trigger method:

a.

Press

Trigger

>

Gated

.

b.

Press

Trigger

>

Trigger Setup

and note that for the

Gate Active Low High

softkey, the default selection is

High, which is the selection used in this example.

6.

Select the trigger source and rear panel input:

a.

For the

Trigger Source

softkey, the default selection is

Ext

, which is the selection used in this

example (gated triggering requires an external trigger).

b.

Press

Trigger Source

and note that for the

Ext Source

softkey, the default selection is

Patt Trig In 1

,

which is the selection used in this example.

7.

Generate the waveform:

Press

ARB Off On

to On.

8.

On the function generator, configure a TTL signal for the external gating trigger.

9.

(Optional) Monitor the current waveform:

Configure the oscilloscope to display both the output of the signal generator, and the external
triggering signal. You will see the waveform modulating the output during the gate inactive
periods (low).

The following figure shows an example display.