Wait timeout – Agilent Technologies Quaternary Pump 1100 Series User Manual
Page 81
![background image](/manuals/33909/81/background.png)
81
Troubleshooting and Test Functions
Wait Timeout
Wait Timeout
When running certain tests in the diagnostics mode or other special
applications, the pump must wait for the plungers to reach a specific
position, or must wait for a certain pressure or flow to be reached. Each
action or state must be completed within the timeout period, otherwise the
error message is generated.
Possible Reasons for a
Wait Timeout
•
Pressure not reached.
•
Pump channel A did not reach the delivery phase.
•
Pump channel B did not reach the delivery phase.
•
Pump channel A did not reach the take-in phase.
•
Pump channel B did not reach the take-in phase.
•
Solvent volume not delivered within the specified time.
Probable Causes
•
Purge valve still open.
•
Leak at fittings, purge valve, active inlet valve, outlet ball valve or plunger
seals.
•
Flow changed after starting test.
•
Defective pump drive assembly.
Suggested Actions
❏ Ensure that purge valve is closed.
❏ Exchange defective capillaries.
❏ Ensure pump components are seated correctly. If there are still signs of a
leak, exchange the appropriate seal (purge valve, active inlet valve, outlet
ball valve, plunger seal).
❏ Ensure correct operating condition for the special application in use.
❏ Exchange the defective pump drive assembly.