Remote cables, Agilent 1100 to 3390 integrators – Agilent Technologies Quaternary Pump 1100 Series User Manual
Page 185
![background image](/manuals/33909/185/background.png)
185
Parts and Materials
Remote Cables
Remote Cables
One end of these cables provides a Agilent Technologies APG (Analytical
Products Group) remote connector to be connected to Agilent 1100 Series
modules. The other end depends on the instrument to be connected to.
Agilent 1100 to 3390 Integrators
Connector
01046-60203
Pin
3390
Pin
Agilent 1100
Signal Name
Active
(TTL)
2
1 - White
Digital ground
NC
2 - Brown
Prepare run
Low
7
3 - Gray
Start
Low
NC
4 - Blue
Shut down
Low
NC
5 - Pink
Not connected
NC
6 - Yellow
Power on
High
NC
7 - Red
Ready
High
NC
8 - Green
Stop
Low
NC
9 - Black
Start request
Low