Agilent Technologies Quaternary Pump 1100 Series User Manual
Page 221
![background image](/manuals/33909/221/background.png)
221
Introduction to the Quaternary Pump
Agilent 1100 Series Interfaces
Table 33
APG Remote Signal Distribution
Pin
Signal
Description
1
DGND
Digital ground
2
PREPARE
(L) Request to prepare for analysis (for example, detector lamp
on). Receiver is any module performing preanalysis activities.
3
START
(L) Request to start run / timetable. Receiver is any module
performing run-time controlled activities.
4
SHUT DOWN
(L) System has serious problem (for example, leak: stops pump).
Receiver is any module that needs to be shutdown to avoid a
safety risk.
5
Not used
6
POWER ON
(H) All modules connected to system are switched on. Receiver is
any module relying on operation of others.
7
READY
(H) System is ready for next analysis. Receiver is any sequence
controller.
8
STOP
(L) Request to reach system ready state as soon as possible (for
example, stop run, abort or finish and stop injection). Receiver is
any module performing run-time controlled activities.
9
START REQUEST (L) Request to start injection cycle. Receiver is the autosampler.