Agilent Technologies Quaternary Pump 1100 Series User Manual
Page 41
![background image](/manuals/33909/41/background.png)
41
Optimizing Performance
Optimize the Compressibility Compensation Setting
5
Starting with a compressibility setting of 10 × 10
-6
/bar increase the value in
steps of 10. Re-zero the integrator as required. The compressibility
compensation setting that generates the smallest pressure ripple is the
optimum value for your solvent composition.
Table 6
Solvent Compressibility
Solvent (pure)
Compressibility (10
-6
/bar)
Acetone
126
Acetonitrile
115
Benzene
95
Carbon tetrachloride
110
Chloroform
100
Cyclohexane
118
Ethanol
114
Ethyl acetate
104
Heptane
120
Hexane
150
Isobutanol
100
Isopropanol 100
Methanol
120
1-Propanol
100
Toluene
87
Water