Agilent Technologies Quaternary Pump 1100 Series User Manual
Page 30
![background image](/manuals/33909/30/background.png)
30
Installing the Pump
Priming and Purging the System
Table 5
Choice of Priming Solvents for Different Purposes
Activity
Solvent
Comments
After an installation
When switching between
reverse phase and normal
phase (both times)
Isopropanol
Isopropanol
Best solvent to flush air out of
the system
Best solvent to flush air out of
the system
After an installation
Ethanol or Methanol
Alternative to Isopropanol
(second choice) if no
Isopropanol is available
To clean the system when
using buffers
After a solvent change
Bidistilled water
Bidistilled water
Best solvent to re-dissolve
buffer cristals
Best solvent to re-dissolve
buffer cristals
After the installation of
normal phase seals (P/N
0905-1420)
Hexane + 5% Isopropanol
Good wetting properties