Lan board, Recommended cables – Agilent Technologies Quaternary Pump 1100 Series User Manual
Page 217
![background image](/manuals/33909/217/background.png)
217
Introduction to the Quaternary Pump
Optional Interface Boards
LAN Board
The HP LAN board is actually an HP JetDirect card, which is a network
interface card used in HP printers.
N O T E
One board is required per Agilent 1100 stack. If the Agilent 1100 stack
has a DAD, then the DAD MUST be the module used for LAN board
installation. If no DAD is present a pump should be used for the LAN
board installation.
N O T E
The LAN board can only be used together with:
A DAD/MWD/VWD/Pump/ALS with main board version G13XX-66520 and
greater. An FLD/RID with main board version G13XX-66500 and greater.
a DOS-ChemStation software revision A.06.01 or above.
The following boards can be used with the Agilent 1100 modules.
N O T E
Minimum firmware of the LAN boards is A.05.05.
Recommended Cables
For point to point connection (not using a network hub) use a twisted pair
cross over LAN cable (P/N 5183-4649, 10 feet long).
For standard network connections using a hub use category 5 UTP cables,
(P/N G1530-61480, 8 m long).
Table 30
LAN Boards
Agilent Order Number Supported networks
J4106A
Ethernet/802.3, RJ-45 (10Base-T)
J4105A
Token Ring/802.5, DB9, RJ-45 (10Base-T)
J4100A
Fast Ethernet, Ethernet/802.3, RJ-45 (10/100Base-TX) + BNC
(10Base2)