Performance specifications – Agilent Technologies Quaternary Pump 1100 Series User Manual
Page 254

254
Specifications
Performance Specifications
Performance Specifications
Table 44
Performance Specification Agilent 1100 Series Quaternary Pump
Type
Specification
Hydraulic system
Dual plunger in series pump with proprietary servo-controlled variable
stroke drive, floating plungers and active inlet valve
Setable flow range
0.001 – 10 ml/min, in 0.001 ml/min increments
Flow range
0.2 – 10.0 ml/min
Flow precision
< 0.3 % RSD (typically 0.15 %), based on retention time, at 1 ml/min
Pressure
Operating range 0 – 40 MPa (0 – 400 bar, 0 – 5880 psi) up to 5 ml/min
Operating range 0 – 20 MPa (0 – 200 bar, 0 – 2950 psi) up to 10 ml/min
Pressure pulsation
< 2 %amplitude (typically < 1 %), at 1 ml/min isopropanol,
at all pressures > 1 MPa (10bar)
Compressibility
compensation
User-selectable, based on mobile phase compressibility
Recommended pH range
1.0 – 12.5, solvents with pH > 2.3 should not contain acids which attack
stainless steel
Gradient formation
Low pressure quaternary mixing/gradient capability using
proprietary high-speed proportioning valve
Delay volume 800 – 1100
µ
l, dependent on back pressure
Composition Range
0 – 95 % or 5 – 100 %, user selectable
Composition Precision < 0.2 % SD, at 0.2 and 1 ml/min
Control and data
evaluation
Agilent ChemStation for LC