Specifications

Performance Specifications

Performance Specifications

Table 44

Performance Specification Agilent 1100 Series Quaternary Pump

Type

Specification

Hydraulic system

Dual plunger in series pump with proprietary servo-controlled variable
stroke drive, floating plungers and active inlet valve

Setable flow range

0.001 – 10 ml/min, in 0.001 ml/min increments

Flow range

0.2 – 10.0 ml/min

Flow precision

< 0.3 % RSD (typically 0.15 %), based on retention time, at 1 ml/min

Pressure

Operating range 0 – 40 MPa (0 – 400 bar, 0 – 5880 psi) up to 5 ml/min

Operating range 0 – 20 MPa (0 – 200 bar, 0 – 2950 psi) up to 10 ml/min

Pressure pulsation

< 2 %amplitude (typically < 1 %), at 1 ml/min isopropanol,
at all pressures > 1 MPa (10bar)

Compressibility
compensation

User-selectable, based on mobile phase compressibility

Recommended pH range

1.0 – 12.5, solvents with pH > 2.3 should not contain acids which attack
stainless steel

Gradient formation

Low pressure quaternary mixing/gradient capability using
proprietary high-speed proportioning valve
Delay volume 800 – 1100

µ

l, dependent on back pressure

Composition Range

0 – 95 % or 5 – 100 %, user selectable

Composition Precision < 0.2 % SD, at 0.2 and 1 ml/min

Control and data
evaluation

Agilent ChemStation for LC